T1 - Overcoming metallic artefacts from orthopaedic wrist volar plating on a low-field MRI scanner
AU - STECCO, ALESSANDRO
AU - Arioli, Roberto
AU - Buemi, Francesco
AU - Parziale, Giuseppe
AU - TRISOGLIO, Alessandra
AU - Soligo, Eleonora
AU - Cerini, Paolo
AU - Leigheb, Massimiliano
AU - Brambilla, Marco
AU - Di Nardo, Gerardo
AU - Guzzardi, Giuseppe
AU - CARRIERO, Alessandro
N1 - Publisher Copyright:
© 2018, Italian Society of Medical Radiology.
PY - 2019
Y1 - 2019
N2 - PURPOSE:
To quantitatively compare the artefact reduction between standard and metallic artefact reduction (MAR) fast spin echo (FSE) T2 sequences in a low-field magnetic resonance imaging (MRI) scanner (0.3 T) in patients with titanium volar wrist plating.
MATERIALS AND METHODS:
Sixteen patients with fractures of the distal radius, treated with titanium volar wrist plating and screws, were examined using a dedicated 0.3 T MRI scanner. Coronal standard FSE T2, FSE T2 high bandwidth (HiBW) and FSE T2 view angle tilting (VAT) sequences were performed. Metallic artefact volume, consisting of both "black" and "bright" artefacts, was calculated for each sequence. Quantitative differences were compared using repeated measures ANOVA test (P < 0.05).
RESULTS:
FSE T2 HiBW and FSE T2 VAT showed a significant reduction in artefact volume compared to the standard sequence. Differences between the artefact volume of the standard FSE T2, HiBW and VAT sequences were statistically significant for both the "black" and "bright" artefacts (P < 0.0001). Differences between the 1.5 HiBW and VAT sequences were statistically significant (black P < 0.0001, bright P < 0.0302).
CONCLUSIONS:
MAR sequences significantly reduced metallic artefacts in vivo using a 0.3 T MRI scanner.
AB - PURPOSE:
To quantitatively compare the artefact reduction between standard and metallic artefact reduction (MAR) fast spin echo (FSE) T2 sequences in a low-field magnetic resonance imaging (MRI) scanner (0.3 T) in patients with titanium volar wrist plating.
MATERIALS AND METHODS:
Sixteen patients with fractures of the distal radius, treated with titanium volar wrist plating and screws, were examined using a dedicated 0.3 T MRI scanner. Coronal standard FSE T2, FSE T2 high bandwidth (HiBW) and FSE T2 view angle tilting (VAT) sequences were performed. Metallic artefact volume, consisting of both "black" and "bright" artefacts, was calculated for each sequence. Quantitative differences were compared using repeated measures ANOVA test (P < 0.05).
RESULTS:
FSE T2 HiBW and FSE T2 VAT showed a significant reduction in artefact volume compared to the standard sequence. Differences between the artefact volume of the standard FSE T2, HiBW and VAT sequences were statistically significant for both the "black" and "bright" artefacts (P < 0.0001). Differences between the 1.5 HiBW and VAT sequences were statistically significant (black P < 0.0001, bright P < 0.0302).
CONCLUSIONS:
MAR sequences significantly reduced metallic artefacts in vivo using a 0.3 T MRI scanner.
KW - Artefacts
KW - MAR
KW - MRI
KW - Plating
KW - Wrist
KW - Artefacts
KW - MAR
KW - MRI
KW - Plating
KW - Wrist
UR - https://iris.uniupo.it/handle/11579/99797
U2 - 10.1007/s11547-018-0973-5
DO - 10.1007/s11547-018-0973-5
M3 - Article
SN - 1826-6983
VL - 124
SP - 392
EP - 399
JO - LA RADIOLOGIA MEDICA
JF - LA RADIOLOGIA MEDICA
IS - 5
ER -