Salta alla navigazione principale Salta alla ricerca Salta al contenuto principale

Operation and calibration of the Silicon Drift Detectors of the ALICE experiment during the 2008 cosmic ray data taking period

  • B. Alessandro
  • , S. Antinori
  • , R. Bala
  • , G. Batigne
  • , S. Beolè
  • , E. Biolcati
  • , N. Bock Garcia
  • , E. Bruna
  • , P. Cerello
  • , S. Coli
  • , Y. Corrales Morales
  • , F. Costa
  • , E. Crescio
  • , P. De Remigis
  • , S. Di Liberto
  • , D. Falchieri
  • , G. Feofilov
  • , W. Ferrarese
  • , E. Gandolfi
  • , C. Garcia
  • L. Gaudichet, G. Giraudo, P. Giubellino, J. Humanic, S. Igolkin, M. Idzik, S. K. Kiprich, A. Kisiel, A. Kolozhvari, I. Kotov, J. Kral, S. Kushpil, V. Kushpil, R. Lea, A. Lisa, I. Martinez, A. Marzari Chiesa, M. Masera, M. Masetti, G. Mazza, A. Mazzoni, F. Meddi, M. Montano Zetina, M. Monteno, S. Nilsen, D. Nouais, F. Padilla Cabal, V. Petráček, G. Poghosyan, F. Prino, L. Ramello, A. Rashevsky, L. Riccati, A. Rivetti, S. Senyukov, M. Siciliano, M. Sitta, A. Subieta Vasquez, M. Sumbera, L. Toscano, F. Tosello, D. Truesdale, M. Urciuoli, A. Vacchi, S. Vallero, A. Werbrouck, G. Zampa, G. Zinovjev

Risultato della ricerca: Contributo su rivistaArticolo in rivistapeer review

Abstract

The calibration and performance of the Silicon Drift Detector of the ALICE experiment during the 2008 cosmic ray run will be presented. In particular the procedures to monitor the running parameters (baselines, noise, drift speed) are detailed. Other relevant parameters (SOP delay, time-zero, charge calibration) were also determined.