N2 - The purpose of this study was to analyze the behavior of a contouring algorithm for PET images based on adaptive thresholding depending on lesions size and target-to-background (TB) ratio under different conditions of image reconstruction parameters. Based on this analysis, the image reconstruction scheme able to maximize the goodness of fit of the thresholding algorithm has been selected. A phantom study employing spherical targets was designed to determine slice-specific threshold (TS) levels which produce accurate cross-sectional areas. A wide range of TB ratio was investigated. Multiple regression methods were used to fit the data and to construct algorithms depending both on target cross-sectional area and TB ratio, using various reconstruction schemes employing a wide range of iteration number and amount of postfiltering Gaussian smoothing. Analysis of covariance was used to test the influence of iteration number and smoothing on threshold determination. The degree of convergence of ordered-subset expectation maximization (OSEM) algorithms does not influence TS determination. Among these approaches, the OSEM at two iterations and eight subsets with a 6-8 mm post-reconstruction Gaussian three-dimensional filter provided the best fit with a coefficient of determination R2 = 0.90 for cross-sectional areas ≤ 133 mm2 and R2 = 0.95 for cross-sectional areas > 133 mm2. The amount of post-reconstruction smoothing has been directly incorporated in the adaptive thresholding algorithms. The feasibility of the method was tested in two patients with lymph node FDG accumulation and in five patients using the bladder to mimic an anatomical structure of large size and uniform uptake, with satisfactory results. Slice-specific adaptive thresholding algorithms look promising as a reproducible method for delineating PET target volumes with good accuracy.
AB - The purpose of this study was to analyze the behavior of a contouring algorithm for PET images based on adaptive thresholding depending on lesions size and target-to-background (TB) ratio under different conditions of image reconstruction parameters. Based on this analysis, the image reconstruction scheme able to maximize the goodness of fit of the thresholding algorithm has been selected. A phantom study employing spherical targets was designed to determine slice-specific threshold (TS) levels which produce accurate cross-sectional areas. A wide range of TB ratio was investigated. Multiple regression methods were used to fit the data and to construct algorithms depending both on target cross-sectional area and TB ratio, using various reconstruction schemes employing a wide range of iteration number and amount of postfiltering Gaussian smoothing. Analysis of covariance was used to test the influence of iteration number and smoothing on threshold determination. The degree of convergence of ordered-subset expectation maximization (OSEM) algorithms does not influence TS determination. Among these approaches, the OSEM at two iterations and eight subsets with a 6-8 mm post-reconstruction Gaussian three-dimensional filter provided the best fit with a coefficient of determination R2 = 0.90 for cross-sectional areas ≤ 133 mm2 and R2 = 0.95 for cross-sectional areas > 133 mm2. The amount of post-reconstruction smoothing has been directly incorporated in the adaptive thresholding algorithms. The feasibility of the method was tested in two patients with lymph node FDG accumulation and in five patients using the bladder to mimic an anatomical structure of large size and uniform uptake, with satisfactory results. Slice-specific adaptive thresholding algorithms look promising as a reproducible method for delineating PET target volumes with good accuracy.
KW - FDG-PET/CT
KW - Functional imaging
KW - Radiation treatment planning
KW - Target volume definition
UR - http://www.scopus.com/inward/record.url?scp=79960535379&partnerID=8YFLogxK
U2 - 10.1120/jacmp.v12i2.3363
DO - 10.1120/jacmp.v12i2.3363
M3 - Article
SN - 1526-9914
VL - 12
SP - 115
EP - 132
JO - Journal of Applied Clinical Medical Physics
JF - Journal of Applied Clinical Medical Physics
IS - 2
ER -