Angio-RM con tecnica "a bolo di contrasto" del rene trapiantato con apparecchiatura a basso campo (0,5 T): Accuratezza diagnostica, sensibilità e specificità delle immagini e delle ricostruzioni nella valutazione delle complicazioni vascolari

Risultato della ricerca: Contributo su rivistaArticolo in rivistapeer review

Abstract

Purpose. This study was performed to assess the feasibility, interobserver variability, sensitivity, specificity and diagnostic accuracy of raw data and postprocessed images from a low-field (0.5-T) magnetic resonance (MR) unit in evaluating vascular complications of kidney grafts. Materials and methods. We enrolled 49 patients undergoing MR angiography (MRA) for a clinical suspicion of renal artery stenosis. The raw data, maximum intensity projections (MIP) and multiplanar reconstruction (MPR) image sets were evaluated independently. We calculated the number and degree of stenosis, and sensitivity, specificity and accuracy for MIP, MPR and raw data image sets, together with interobserver variability. Results. Interobserver agreement was substantial. There were no differences among the MIP and MPR algorithms and raw data images for the detection of stenosis. Raw data images were more accurate in quantifying the severity of stenosis, with higher sensitivity (75% vs. 62.5%), equal specificity and higher diagnostic accuracy (75% vs. 71.43%). Conclusions. Contrast-enhanced MRA, even with a low-field (0.5-T) unit, is a feasible, sensitive and accurate technique for the study of the renal arteries of the transplanted kidney.

Titolo tradotto del contributoContrast-bolus MR angiography of the transplanted kidney with a low-field (0.5-T) scanner: Diagnostic accuracy, sensitivity and specificity of images and reconstructions in the evaluation of vascular complications
Lingua originaleItalian
pagine (da-a)1026-1035
Numero di pagine10
RivistaRadiologia Medica
Volume112
Numero di pubblicazione7
DOI
Stato di pubblicazionePubblicato - ott 2007

Keywords

  • Angio-MRI
  • Renal Transplant
  • Renal artery stenosis

Fingerprint

Entra nei temi di ricerca di 'Angio-RM con tecnica "a bolo di contrasto" del rene trapiantato con apparecchiatura a basso campo (0,5 T): Accuratezza diagnostica, sensibilità e specificità delle immagini e delle ricostruzioni nella valutazione delle complicazioni vascolari'. Insieme formano una fingerprint unica.

Cita questo