Abstract
A high-pressure liquid chromatograph equipped with an off-column calorimetric reactor was used for the determination of metallorganic compounds such as metallothioneins. The stainless-steel gel-permeation column used was first tested with ultraviolet detection of a standard protein mixture. The metallothioneins were determined by absorbance measurements at 520 nm after an off-column reaction of the metal-containing eluates with a buffered 4-(2-pyridylazo)resorcinol (PAR) solution. The eluates were also tested by flame and electrothermal atomic-absorption spectrometry; a correlation between AAS and ICP results and the peak areas of the calorimetric detection was also made. A zinc-inhibiting reagent was also added to the PAR solution in order to find the calorimetric error due to the presence of other metals such as copper or cadmium.
| Original language | English |
|---|---|
| Pages (from-to) | 693-696 |
| Number of pages | 4 |
| Journal | Talanta |
| Volume | 35 |
| Issue number | 9 |
| DOIs | |
| Publication status | Published - Sept 1988 |
| Externally published | Yes |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 14 Life Below Water
Fingerprint
Dive into the research topics of 'The speciation of trace amounts of organometallics in marine organisms by gel-permeation high-pressure liquid chromatography with PAR derivatization'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver