Skip to main navigation Skip to search Skip to main content

Osseointegration of sandblasted or anodised hydrothermally-treated titanium implants: Mechanical, histomorphometric and bone hardness measurements

  • G. Giavaresi
  • , M. Fini
  • , R. Chiesa
  • , L. Rimondini
  • , G. Rondelli
  • , V. Borsari
  • , L. Martini
  • , N. Nicoli Aldini
  • , G. A. Guzzardella
  • , Roberto Giardino

Research output: Contribution to journalArticlepeer-review

Abstract

The improvement of the implant-bone interface is still an open problem in the long-term mechanical stability of cementless fixed implants. Mechanical, histomorphometric and bone hardness measurements were performed in sheep femoral cortical bone implants at 8 and 12 weeks from surgery to compare in vivo the osseointegration of titanium screws (Ø 3.5 mm × 7 mm length) with two different surface treatments: sandblasting with 70-100 μm HA followed by acid etching with HNO3 (Group A) and Ca-P anodization followed by a hydrothermal treatment (Group B). No significant differences were found for maximum push-out force and interfacial strength between groups at both experimental times. No significant difference was observed for Bone Ingrowth between groups at both experimental times, while the Affinity Index of Group B was significantly higher (7.5%, p<0.05) and lower (10.2%, p<0.05) than that of Group A at 8 and 12 weeks, respectively. Finally, a significant increase in bone microhardness measured within 200 μm from the interface and inside the thread depth of Group A was observed between the two experimental times (p<0.05). In conclusion, present findings show that osseointegration may be accelerated by adequate surface roughness and bioactive ceramic coating such as current tested treatments which enhance bone interlocking and mineralization.

Original languageEnglish
Pages (from-to)806-813
Number of pages8
JournalInternational Journal of Artificial Organs
Volume25
Issue number8
DOIs
Publication statusPublished - 1 Aug 2002
Externally publishedYes

Keywords

  • Bone histomorphometry
  • Bone microhardness
  • Electrochemical surface treatment
  • Interfacial strength
  • Sandblasted and acid-etched surface
  • Titanium implants

Fingerprint

Dive into the research topics of 'Osseointegration of sandblasted or anodised hydrothermally-treated titanium implants: Mechanical, histomorphometric and bone hardness measurements'. Together they form a unique fingerprint.

Cite this