Abstract
The paper discusses a class of stochastic models for evaluating the optimal calibration interval in measuring instruments. The model is based on the assumption that the calibration status of a measuring instrument can be monitored by means of one observable parameter. The observable parameter is undergoing a stochastic drift process. The paper introduces and compares stochastic drift models of different nature, and estimates the first passage time of the monitored parameter on a preset limit. The calibration interval is determined as a suitable percentile of the distribution function of the first passage time. A preliminary validation of the model, based on a sample of experimental data collected on a class of instruments, is finally reported.
| Original language | English |
|---|---|
| Pages (from-to) | 747-751 |
| Number of pages | 5 |
| Journal | IEEE Transactions on Instrumentation and Measurement |
| Volume | 46 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - 1997 |
| Externally published | Yes |
Keywords
- Calibration intervals
- Measuring instrument
- Stochastic models
Fingerprint