Influence of ventilator settings on patient- ventilator synchrony during pressure support ventilation with different interfaces

  • R. Costa
  • , P. Navalesi
  • , G. Spinazzola
  • , G. Ferrone
  • , A. Pellegrini
  • , F. Cavaliere
  • , R. Proietti
  • , M. Antonelli
  • , G. Conti

Research output: Contribution to journalArticlepeer-review

Abstract

Objective: To evaluate patient-ventilator interaction during pressure support ventilation (PSV) delivered with three interfaces [endotracheal tube (ET), face mask (FM), and helmet (H)] at different pressurization times (Timepress), cycling-off flow thresholds (Trexp), and respiratory rates (RR) in a bench study, and with FM and H in a healthy volunteers study. Design: Bench study using a mannequin connected to an active lung simulator, and human study including eight healthy volunteers. Measurements: PSV was delivered through the three interfaces with three different RR in the bench study, and through FM and H at two different RR in the human study. The mechanical and the neural RR, Ti, Te, inspiratory trigger delay (Delaytrinsp), pressurization time, and expiratory trigger delay were randomly evaluated at various ventilator settings (Timepress/Trexp: 50%/25%, default setting; 20%/5%, slow setting; 80%/60%, fast setting). Results: Bench study: patient-ventilator synchrony was significantly better with ET, with lower Delaytrinsp and higher time of assistance (P<0.001); the combination Timepress/Trexp 20%/5% at RR 30 produced the worst interaction, with higher rate of wasted efforts (WE) compared with Time press/Trexp 80%/60% (20%, 40%, and 50% of WE versus 0%, 16%, and 26% of all spontaneous breaths, with ET, FM, and H, respectively; P<0.01). In both studies, compared with H, FM resulted in better synchrony. Conclusion: Patient-ventilator synchrony was significantly better with ET during the bench study; in the human study, FM outperformed H.

Original languageEnglish
Pages (from-to)1363-1370
Number of pages8
JournalIntensive Care Medicine
Volume36
Issue number8
DOIs
Publication statusPublished - Aug 2010
Externally publishedYes

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Endotracheal tube
  • Face mask
  • Helmet
  • Noninvasive ventilation
  • Patient-ventilator interaction
  • Pressurization rate
  • Trigger

Fingerprint

Dive into the research topics of 'Influence of ventilator settings on patient- ventilator synchrony during pressure support ventilation with different interfaces'. Together they form a unique fingerprint.

Cite this