Skip to main navigation Skip to search Skip to main content

Heart rate assessment using NeoTapAdvancedSupport: A simulation study

  • Marco Binotti
  • , Francesco Cavallin
  • , Pier Luigi Ingrassia
  • , Nicolas J. Pejovic
  • , Alice Monzani
  • , Giulia Genoni
  • , Daniele Trevisanuto

Research output: Contribution to journalArticlepeer-review

Abstract

Background NeoTapAdvancedSupport (NeoTapAS) is a mobile application, based on a screen tapping method that calculates the heart rate (HR). We aimed to evaluate the accuracy of NeoTapAS in reliably determining HR from auscultation in a high-fidelity simulated newborn resuscitation scenario. Methods Paediatric residents assessed HR by auscultation plus NeoTapAS in an asphyxiated term infant scenario and orally communicated the estimated HR. An external observer simultaneously documented the actual HR set in the manikin and the communicated HR. Results One hundred and sixty HR measurements were recorded. The agreement between communicated and set HR was good (Cohen's kappa 0.80, 95% CI 0.72 to 0.87; Bangdiwala's weighted agreement strength statistic 0.93). Bland-Altman plot showed a mean difference between communicated and set HR values of 1 beats per minute (bpm) (95% agreement limits -9 to 11 bpm). Conclusion NeoTapAS showed a good accuracy in estimating HR and it could be an important resource in settings with limited availability of ECG monitor.

Original languageEnglish
Pages (from-to)F440-F442
JournalArchives of Disease in Childhood: Fetal and Neonatal Edition
Volume104
Issue number4
DOIs
Publication statusPublished - 1 Jul 2019

Keywords

  • accuracy
  • heart rate
  • mobile health
  • neonatology
  • resuscitation

Fingerprint

Dive into the research topics of 'Heart rate assessment using NeoTapAdvancedSupport: A simulation study'. Together they form a unique fingerprint.

Cite this